How to deal with the problem of laboratory ultra-pure water machine? In response to such a situation, Shanghai Bayu Industrial Co., Ltd. invited the technician Su Shifu to answer the detailed troubleshooting of the laboratory ultra-pure water machine!
1. The quality of the primary effluent is poor
1. It is possible to deposit water for too long and drain it.
2. The circulation pump may be damaged, it is not possible to recycle, and the circulation pump is replaced.
3. It may be that the (super)purification column fails and the (super)purification column is replaced.
4. The Check Valve on the circulation line may be damaged, causing the water not to pass through the (super)purification column and directly return the water through the circulation line, and replace the check valve on the circulation line.
5. The system function is abnormal: it may be that the system circuit is damaged and the circuit board is replaced.
2. The system does not produce water
1. Shorting may cause the booster pump to idle and damage the pump.
2. The low voltage protection switch is damaged, it cannot be closed normally, and the low voltage protection switch is replaced.
3. It may be that the pre-treatment Filter element is clogged, the water flux is reduced or no water passes, and the pre-treatment filter element is replaced.
4. If the tap water is not open or the water pressure is not enough, the system will sound an alarm, turn on the tap water or short-circuit the low-voltage protection switch.
5. The high voltage protection switch is damaged and cannot be closed normally. At this time, the full water indicator will light up and replace the high voltage protection switch.
6. System water and full water conversion are too frequent: it may be that the pure water bucket Ball Valve is not open, can not store water, open the pure water bucket ball valve.
7. The system has always made water: the high-voltage protection switch is damaged, it can't be disconnected normally, the water-making indicator light is always on, and the high-voltage protection switch is replaced.
3. System water production is small
1. The reverse osmosis membrane is clogged and the reverse osmosis membrane is replaced.
2. Insufficient water supply, improved water supply or replacement of pre-treatment filter elements.
3. It may be that the booster pump is damaged and replace the booster pump.
4. One (three) level of water intake is small
1. There may be too little water in the pure water tank, and the water will be taken after the system is full of water.
2. It may be that the pressure in the pure water tank airbag is too low or no pressure, and the airbag of the pure water tank is reinflated.
The above is how to deal with the problem of laboratory ultra-pure water machine failure? For the full content of the book, if you want to know more about the laboratory ultra-pure water machine or want to buy other laboratory equipment products, you can contact Shanghai Bashu Industrial Co., Ltd. at any time!